GSO IEC 62127-2:2026
GSO - Cover page
Gulf Foreword
IEC - Cover page
FOREWORD
INTRODUCTION
1 Scope
2 Normative references
3 Terms and definitions
4 List of symbols
5 Overview of calibration procedures
 5.1 Principles
 5.2 Summary of calibration procedures
 5.3 Reporting of results
 5.4 Recommended calibration periods
6 Generic requirements of a hydrophone calibration system
 6.1 Mechanical positioning
  6.1.1 General
  6.1.2 Accuracy of the axial hydrophone position
  6.1.3 Accuracy of the lateral hydrophone position
 6.2 Temperature measurements and temperature stability
 6.3 Hydrophone size
 6.4 Measurement vessel and water properties
 6.5 Measurement of output voltage
7 Electrical considerations
 7.1 Signal type
 7.2 Earthing
 7.3 Measurement of hydrophone output voltage
  7.3.1 General
  7.3.2 Electrical loading by measuring instrument
  7.3.3 Electrical loading by extension cables
  7.3.4 Noise
  7.3.5 Cross-talk (radio-frequency rf pick-up) and acoustic interference
  7.3.6 Integral hydrophone pre-amplifiers
8 Preparation of hydrophones
 8.1 General
 8.2 Wetting
 8.3 Hydrophone support
 8.4 Influence of cable
9 Free field reciprocity calibration
 9.1 General
 9.2 Object
 9.3 General principles
  9.3.1 General
  9.3.2 Three-transducer reciprocity calibration method
  9.3.3 Self-reciprocity calibration method
  9.3.4 Two-transducer reciprocity calibration method
 9.4 Two-transducer reciprocity calibration method
  9.4.1 General
  9.4.2 Auxiliary transducers
  9.4.3 Reflector
  9.4.4 Measurement field
  9.4.5 Reciprocity approach
  9.4.6 Measurement procedure
10 Free field calibration by planar scanning
 10.1 General
 10.2 Object
 10.3 General principle
 10.4 Procedural requirements
  10.4.1 Hydrophone scanning
  10.4.2 Power measurement
  10.4.3 Transducer mounting
  10.4.4 Measurement conditions
  10.4.5 Measurements
 10.5 Corrections and sources of uncertainty
11 Free field calibration by optical interferometry
 11.1 General
 11.2 Principle
12 Calibration by comparison using a standard hydrophone
 12.1 General
 12.2 Object
 12.3 Principle
 12.4 Procedural requirements
  12.4.1 Source transducer
  12.4.2 Source transducer drive signal
  12.4.3 Measurement system
 12.5 Procedure
  12.5.1 Measurements (Type I): determination of the directional response of a hydrophone
  12.5.2 Measurements (Type II): calibration by comparison using a standard hydrophone
 12.6 Maximum hydrophone size
Annex A (informative) Assessment of uncertainty in free field calibration of hydrophones
Annex B (informative) Behaviour of PVDF polymer sensors in high-intensity ultrasonic fields
Annex C (informative) Electrical loading corrections
Annex D (informative) Absolute calibration of hydrophones using the planar scanning technique
Annex E (informative) Properties of water
Annex F (informative) The absolute calibration of hydrophones by optical interferometry
Annex G (informative) Waveform concepts
Annex H (informative) Time delay spectrometry - Requirements and a brief review of the technique
Annex I (informative) Determination of the phase response of hydrophones
Annex J (informative) Maximum size considerations for the active element of a hydrophone
Annex K (informative) Two-transducer reciprocity calibration method
Bibliography
30mm
20mm
15mm
15mm
10mm
17mm
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard
GSO IEC 62127-2:2026
IEC 62127-2:2025
مواصفة قياسية خليجية
الموجات فوق الصوتية-السماعات المائية - (الهيدروفونات) - الجزء 2: معايرة مجالات الموجات فوق الصوتية
Gulf Standard
Ultrasonics - Hydrophones - Part 2: Calibration for ultrasonic fields
ICS:
17.140.50

svg+xml;base64,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
هذه الوثيقة محمية بالقانون
COPYRIGHT PROTECTED DOCUMENT
جميع الحقوق محفوظة لهيئة لتقييس لدول مجلس التعاون لدول الخليج العربية.
ما لم ينص على خلاف ذلك، لا يجوز إعادة إنتاج أي جزء من هذا المنشور أو استخدامه بأي شكل آخر أو بأي وسيلة، إلكترونية أو ميكانيكية، بما في ذلك التصوير، أو النشر على الإنترنت أو الإنترانت، دون إذن كتابي مسبق من هيئة التقييس لدول مجلس التعاون لدول الخليج العربية على العنوان أدناه أو الدول الأعضاء في الهيئة داخل بلدانهم.
All rights reserved to GCC Standardization Organization (GSO). Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either GSO at the address below or GSO's member body in the country of the requester.
Outreach Dept, Knowledge Management Section
GCC Standardization Organization (GSO)
P.O.Box: 85245 Riyadh 12511
Kingdom of Saudi Arabia
إدارة التواصل، قسم إدارة المعرفة
هيئة التقييس لدول مجلس التعاون لدول الخليج العربية (GSO)
ص. ب: 85245 الرياض 12511
المملكة العربية السعودية
Tel: +966115208000 :هاتف
Fax: +966115208010 :فاكس
E-Mail: csc@gso.org.sa :بريد إلكتروني
Website: www.gso.org.sa :موقع ويب
30mm
20mm
15mm
15mm
10mm
17mm

Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard

Gulf Foreword

GCC Standardization Organization (GSO) is a regional organization which consists of the national standardization bodies of GCC member states. One of GSO main functions is to issue gulf standards and technical regulations through specialized technical committees.
GSO Technical Council in its meeting number (62) held on 26/10/1447 H, which corresponds to 14/04/2026, has approved Gulf Standard GSO IEC 62127-2:2026 titled "Ultrasonics - Hydrophones - Part 2: Calibration for ultrasonic fields" in English language which was studied and adopted from International Electrotechnical Commission Standard IEC 62127-2:2025 without any technical modifications through the technical program of GSO Technical Committee for Metrology (TC04) in QATAR schedule .
تقديم خليجي
هيئة التقييس لدول مجلس التعاون لدول الخليج العربية هيئة إقليمية تضم في عضويتها أجهزة التقييس الوطنية في الدول الأعضاء ، ومن مهام الهيئة إعداد المواصفات القياسية واللوائح الفنية الخليجية بواسطة لجان فنية متخصصة.
قرر المجلس الفني لهيئة التقييس لدول مجلس التعاون لدول الخليج العربية في الاجتماع الـ (62) بتاريخ 1447/10/26 هـ ،الموافق 2026/04/14م اعتماد المواصفة القياسية الخليجية رقم GSO IEC 62127-2:2026 وعنوانها "الموجات فوق الصوتية-السماعات المائية - (الهيدروفونات) - الجزء 2: معايرة مجالات الموجات فوق الصوتية" باللغة الانجليزية التي تم دراستها وتبنيها عن اللجنة الدولية الكهروتقنية رقم IEC 62127-2:2025 بدون إدخال تعديلات فنية عليها ضمن برنامج عمل اللجنة الفنية الخليجية لمواصفات المقاييس (TC04) المدرجة في خطة دولة قطر .
30mm
20mm
15mm
15mm
10mm
17mm
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard
IEC 62127-2

INTERNATIONAL

STANDARD

Edition 2.0  2025-01

Ultrasonics –
Hydrophones – Part 2: Calibration for ultrasonic fields

INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


ICS 17.140.50ISBN 9782832700914
Warning! Make sure that you obtained this publication from an authorized distributor.

© Registered trademark of the International Electrotechnical Commission

30mm
20mm
15mm
15mm
10mm
17mm
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard

About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform
The advanced search enables to find IEC publications by a variety of criteria (reference number, text, technical committee, ...). It also gives information on projects, replaced and withdrawn publications.

IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published details all new publications released. Available online and once a month by email.

IEC Customer Service Centre - webstore.iec.ch/csc
If you wish to give us your feedback on this publication or need further assistance, please contact the Customer Service Centre: sales@iec.ch .

 

IEC Products & Services Portal - products.iec.ch
Discover our powerful search engine and read freely all the publications previews. With a subscription you will always have access to up to date content tailored to your needs.

Electropedia - www.electropedia.org
The world's leading online dictionary on electrotechnology, containing more than 22 300 terminological entries in English and French, with equivalent terms in 19 additional languages. Also known as the International Electrotechnical Vocabulary (IEV) online.

30mm
20mm
15mm
15mm
10mm
17mm

Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard
CONTENTS
INTERNATIONAL ELECTROTECHNICAL COMMISSION

International StandardIEC 62127-2:2025 ED2
Ultrasonics - Hydrophones - Part 2: Calibration for ultrasonic fields

FOREWORD

1)
The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC Publication(s)"). Their preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with may participate in this preparatory work. International, governmental and non-governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2)
The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international consensus of opinion on the relevant subjects since each technical committee has representation from all interested IEC National Committees.
3)
IEC Publications have the form of recommendations for international use and are accepted by IEC National Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any misinterpretation by any end user.
4)
In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications transparently to the maximum extent possible in their national and regional publications. Any divergence between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5)
IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any services carried out by independent certification bodies.
6)
All users should ensure that they have the latest edition of this publication.
7)
No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and members of its technical committees and IEC National Committees for any personal injury, property damage or other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC Publications.
8)
Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is indispensable for the correct application of this publication.
9)
IEC draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC shall not be held responsible for identifying any or all such patent rights.
IEC 62127-2 has been prepared by IEC technical committee 87: Ultrasonics. It is an International Standard.
This second edition cancels and replaces the first edition published in 2007, Amendment 1:2013 and Amendment 2:2017. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a)
the upper frequency limit of 40 MHz has been removed;
b)
hydrophone sensitivity definitions have been changed to recognize sensitivities as complex-valued quantities;
c)
directional response measurement and effective size determination procedures have been updated in 12.5.1 to align with recent changes in IEC 62127-3;
d)
Annex F has been amended to comprise a calibration technique for high-frequency complex-valued calibration;
e)
the reciprocity method description in Annex K was extended to also comprise focusing transducers;
The text of this International Standard is based on the following documents:
Draft Report on voting
87/878/FDIS 87/884/RVD
Full information on the voting for its approval can be found in the report on voting indicated in the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are described in greater detail at www.iec.ch/publications.
A list of all parts of IEC 62127 series, published under the general title Ultrasonics - Hydrophones, can be found on the IEC website.
NOTETerms in bold in the text are defined in Clause 3.
The committee has decided that the contents of this document will remain unchanged until the stability date indicated on the IEC website under webstore.iec.ch in the data related to the specific document. At this date, the document will be
•
reconfirmed,
•
withdrawn, or
•
revised.
IMPORTANT
The "colour inside" logo on the cover page of this document indicates that it contains colours which are considered to be useful for the correct understanding of its contents. Users should therefore print this document using a colour printer.

INTRODUCTION

The spatial and temporal distribution of acoustic pressure in an ultrasonic field in a liquid medium is commonly determined using miniature ultrasonic hydrophones. These devices are not absolute measurement instruments and it is important that they are calibrated. This part of IEC 62127 specifies the calibration methods to use in determining the response of a hydrophone in the ultrasonic range, i.e. above 50 kHz. The main hydrophone application in this context lies in the measurement of ultrasonic fields emitted by medical diagnostic equipment in water. It is important to understand hydrophone behaviour over a wide frequency band in order to reliably characterize the acoustic parameters of the applied acoustic field. In particular, the frequency range above 15 MHz is important to fully characterize this equipment, primarily due to the increased appearance of high-frequency components in the ultrasonic signals, caused by nonlinear propagation. In addition, the number of medical ultrasonic systems that use frequencies above 15 MHz, particularly intra-operative probes, is growing. It has turned out in recent years that the hydrophone response below 0,5 MHz is also important in order to reliably determine the peak-negative (rarefactional) acoustic pressure.
While the term "hydrophone" can be used in a wider sense, it is understood here as referring to miniature piezoelectric hydrophones. It is this instrument type that is used today in various areas of medical ultrasonics and, in particular, to characterize quantitatively the field structure of medical diagnostic instruments [1]1 . With regard to other pressure sensor types, such as those based on fibre optics, some of the requirements of this document are applicable to these as well but others are not. If in the future these other "hydrophone" types gain more importance in field measurement practice, their characteristics and calibration will be dealt with in a future edition of IEC 62127-2 or in a separate part of IEC 62127.
NOTE 1This document covers the ultrasonic frequency range, from 50 kHz to an upper frequency of 100 MHz. Not all techniques described are applicable to the full frequency range. Standards dealing with hydrophone properties (IEC 62127-3) and hydrophone use (IEC 62127-1) are being maintained in parallel. This will eventually lead to unified standards covering the whole field of practical hydrophone application.
NOTE 2Hydrophone calibration in the lower ultrasonic and in the underwater sound frequency range is particularly addressed in the IEC 60565 series [2],[3].
30mm
20mm
15mm
15mm
10mm
17mm

Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard
International StandardIEC 62127-2:2025 ED2
Ultrasonics - Hydrophones - Part 2: Calibration for ultrasonic fields

1Scope

This part of IEC 62127 specifies:
•
absolute hydrophone calibration methods;
•
relative (comparative) hydrophone calibration methods.
Recommendations and references to accepted literature are made for the various relative and absolute calibration methods in the frequency range covered by this document.
This document is applicable to
•
hydrophones used for measurements made in water and in the ultrasonic frequency range 50 kHz to 100 MHz;
NOTE 1Although some physiotherapy medical applications of medical ultrasound are developing which operate in the frequency range 40 kHz to 100 kHz, the primary frequency range of diagnostic imaging remains above 2 MHz. It has recently been established that, even in the latter case, the hydrophone response at substantially lower frequencies can influence measurements made of key acoustic parameters [4].
NOTE 2Calibration methods for underwater acoustics hydrophones applicable in the frequency range from 200 Hz to 1 MHz are available in IEC 60565-1 [2], and for frequencies from 0,01 Hz to several kilohertz in IEC 60565-2 [3].
•
hydrophones employing piezoelectric sensor elements, designed to measure the pulsed wave and continuous wave ultrasonic fields generated by ultrasonic equipment;
NOTE 3Some hydrophones can have non-circular active elements, arising from slight deviations from a circular structure caused, for example, by electrode structure; or, conversely, the active elements can actually be squares. It is important in these cases to pay special attention to the directional response and to the effective radii of the active element through various axes of rotation.
•
hydrophones with or without a hydrophone pre-amplifier.

2Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
IEC 61161, Ultrasonics - Power measurement - Radiation force balances and performance requirements
IEC 61689, Ultrasonics - Physiotherapy systems - Field specifications and methods of measurement in the frequency range 0,5 MHz to 5 MHz
IEC 62127-1, Ultrasonics - Hydrophones - Part 1: Measurement and characterization of medical ultrasonic fields
IEC 62127-3:2022, Ultrasonics - Hydrophones - Part 3: Properties of hydrophones for ultrasonic fields

3Terms and definitions

For the purposes of this document, the terms and definitions given in IEC 62127-1, IEC 62127-3 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
•
IEC Electropedia: available at https://www.electropedia.org/
•
ISO Online browsing platform: available at https://www.iso.org/obp
3.1
acoustic centre
point on or near a transducer from which the spherically divergent sound waves emitted by the transducer, and observable at remote points, appear to diverge
Note 1 to entry: This point is the boundary between the near field and the far field.
Note 2 to entry: In this region, the sound field has plane wave fronts changing to spherical wave fronts with divergence.
3.2
beam axis
straight line that passes through the beam centrepoints of two planes perpendicular to the line which connects the point of maximal pulse-pressure-squared integral with the centre of the external transducer surface
Note 1 to entry: The location of the first plane is the location of the plane containing the maximum pulse-pressure-squared integral or, alternatively, is one containing a single main lobe which is in the focal Fraunhofer zone. The location of the second plane is as far as is practicable from the first plane and parallel to the first with the same two orthogonal scan lines (x and y axes) used for the first plane.
Note 2 to entry: In a number of cases, the term pulse-pressure-squared integral is replaced in the above definition by any linearly related quantity, for example
a)
in the case of a continuous wave signal the term pulse-pressure-squared integral is replaced by mean square acoustic pressure as defined in IEC 61689,
b)
in cases where signal synchronization with the scanframe is not available the term pulse-pressure-squared integral may be replaced by temporal average intensity.
Note 3 to entry: See IEC 62127-1:2022, Figure 1.
[SOURCE: IEC 62127-1:2022, 3.8, modified - In the definition, "aperture" has been replaced with "surface".]
3.3
beam centrepoint
position determined by the intersection of two lines in the same beam area plane xy passing through the beamwidth midpoints of two orthogonal planes, xz and yz, perpendicular to their respective beamwidth lines
[SOURCE: IEC 62127-1:2022, 3.9]
3.4
beamwidth
w6, w12, w20
greatest distance between two points on a specified axis perpendicular to the beam axis where the pulse-pressure-squared integral falls below its maximum on the specified axis by a specified amount
Note 1 to entry: In a number of cases, the term pulse-pressure-squared integral is replaced in the above definition by any linearly related quantity, for example
a)
in the case of a continuous wave signal the term pulse-pressure-squared integral is replaced by mean square acoustic pressure as defined in IEC 61689,
b)
in cases where signal synchronization with the scanframe is not available the term pulse-pressure-squared integral may be replaced by temporal average intensity.
Note 2 to entry: Commonly used beamwidths are specified at −6 dB, −12 dB and −20 dB levels below the maximum. The decibel calculation implies taking 10 times the logarithm of the ratios of the integrals.
Note 3 to entry: Beamwidth is expressed in metres (m).
[SOURCE: IEC 62127-1:2022, 3.11]
3.5
beamwidth midpoint
linear average of the location of the centres of beamwidths in a plane
Note 1 to entry: The average is taken over as many beamwidth levels as given in IEC 61828:2020, Clause C.2 and Table C.2.
[SOURCE: IEC 61828:2020 [5], 3.22]
3.6
beam centre
point in a plane in the far field, usually perpendicular to the beam axis, at which the spatial-peak temporal-peak acoustic pressure occurs
3.7
derived instantaneous intensity
quotient of squared instantaneous acoustic pressure and characteristic acoustic impedance of the medium at a particular instant in time at a particular point in an acoustic field
mml-m1 I ( t ) = p ( t ) 2 ρ   c
 (1)
where:
p(t)
ρ
is the density of the medium;
c
is the speed of sound in the medium.
Note 1 to entry: For measurement purposes referred to in this document, the derived instantaneous intensity is an approximation of the instantaneous intensity.
Note 2 to entry: Increased uncertainty is expected to be taken into account for measurements very close to the transducer.
Note 3 to entry: Derived instantaneous intensity is expressed in units of watt per metre squared (W/m2).
[SOURCE: IEC 62127-1:2022, 3.14]
3.8
diametrical beam scan
set of measurements of the hydrophone output voltage made while moving the hydrophone in a straight line passing through a point on the beam axis and in a direction normal to the beam axis
Note 1 to entry: The diametrical beam scan can extend to different distances on either side of the beam axis.
[SOURCE: IEC 62127-1:2022, 3.15]
3.9
directional response
description of the response of a hydrophone, as a function of direction of propagation of the incident plane sound wave, in a specified plane through the reference centre and at a specified frequency
Note 1 to entry: Although directional response is a complex-valued function, it is generally the modulus of directional response that is of most interest and this is commonly presented graphically.
[SOURCE: IEC 62127-3:2022, 3.2]
3.10
effective hydrophone size
ah
size of a theoretical receiver hydrophone that has a predicted directional response function with an angular width equal to the observed angular width
Note 1 to entry: The size is usually the function of frequency. For representative experimental data, see [9].
Note 2 to entry: The effective hydrophone size is expressed in metres (m).
Note 3 to entry: For hydrophones with a circular geometry, the effective hydrophone size is a radius.
Note 4 to entry: For hydrophones with a rectangular geometry, the effective hydrophone size is half the greatest value of the length or width.
[SOURCE: IEC 62127-1:2022, 3.20]
3.11
effective radius of a non-focusing ultrasonic transducer
at
radius of a perfect disc piston-like ultrasonic source transducer that has a predicted axial acoustic pressure distribution approximately equivalent to the observed axial acoustic pressure distribution over an axial distance until at least the last axial maximum has passed
Note 1 to entry: The effective radius of a non-focusing ultrasonic transducer is expressed in metres (m).
[SOURCE: IEC 62127-1:2022, 3.21]
3.12
effective wavelength
λeff
longitudinal speed of sound in the propagation medium divided by the acoustic-working frequency
Note 1 to entry: The effective wavelength is expressed in metres (m).
Note 2 to entry: : Application dependent definitions of the acoustic-working frequency are given in IEC 62127-1:2022.
[SOURCE: IEC 61828:2020 [5], 3.46, modified - Note 2 to entry has been added.]
3.13
electric load impedance
ẔL
complex electric input impedance (consisting of a real and an imaginary part) to which the hydrophone unit output cable is connected or is intended to be connected
Note 1 to entry: The electric load impedance is expressed in ohms (Ω).
[SOURCE: IEC 62127-1:2022, 3.22]
3.14
end-of-cable loaded sensitivity
mml-m3 M _ L ( f )
<of a hydrophone or hydrophone assembly> quotient of the Fourier transformed hydrophone voltage-time signal F (uL (t )) at the end of any integral cable or output connector of a hydrophone or hydrophone-assembly, when connected to a specified electric load impedance, to the Fourier transformed acoustic pulse waveform F (p (t)) in the undisturbed free field of a plane wave in the position of the reference centre of the hydrophone if the hydrophone were removed
mml-m2 M _ L ( f ) = F ( u L ( t ) ) F ( p ( t ) )
 (2)
Note 1 to entry: The end-of-cable loaded sensitivity is a complex-valued parameter. Its modulus is expressed in units of volt per pascal (V/Pa), its phase angle is expressed in degrees, and represents the phase difference between the electrical voltage and the sound pressure.
[SOURCE: IEC 62127-1:2022, 3.25]
3.15
end-of-cable loaded sensitivity level
LML(f)
<of a hydrophone or hydrophone assembly> twenty times the logarithm to the base 10 of the quotient of the modulus of the end-of-cable loaded sensitivity mml-m4 | M _ L ( f ) | to a reference sensitivity Mref
mml-m5 L M L ( f ) = 20   log 10 | M _ L ( f ) | M ref dB
 (3)
Note 1 to entry: Commonly used values of the reference sensitivity Mref are 1 V/µPa or 1 V/Pa.
Note 2 to entry: The end-of-cable loaded sensitivity level is expressed in decibels (dB).
[SOURCE: IEC 62127-1:2022, 3.26]
3.16
end-of-cable open-circuit sensitivity
mml-m7 M _ c ( f )
<of a hydrophone> quotient of the Fourier transformed hydrophone open-circuit voltage-time signal ℱ (uc (t)) at the end of any integral cable or output connector of a hydrophone to the Fourier transformed acoustic pulse waveform ℱ (p (t)) in the undisturbed free field of a plane wave in the position of the reference centre of the hydrophone if the hydrophone were removed
mml-m6 M _ c ( f ) = F ( u c ( t ) ) F ( p ( t ) )
 (4)
Note 1 to entry: The end-of-cable open-circuit sensitivity is a complex-valued parameter. Its modulus is expressed in units of volt per pascal (V/Pa), its phase angle is expressed in degrees, and represents the phase difference between the electrical voltage and the sound pressure.
[SOURCE: IEC 62127-1:2022, 3.27]
3.17
external transducer surface
external transducer aperture
part of the surface of the ultrasonic transducer or ultrasonic transducer element group assembly that emits ultrasonic radiation into the propagation medium
Note 1 to entry: This surface is assumed to be accessible for measurements using a hydrophone in a chosen propagation medium (usually water).
Note 2 to entry: This surface is either directly in contact with the patient or is in contact with a water or liquid path to the patient.
Note 3 to entry: See IEC 61828:2020 [5], Figure B.5.
[SOURCE: IEC 61828:2020 [5], 3.52]
3.18
far field
region of the field where mml-m8 z > z T aligned along the beam axis for planar non-focusing transducers
Note 1 to entry: In the far field, the sound pressure appears to be spherically divergent from a point on or near the radiating surface. Hence, the pressure produced by the sound source is approximately inversely proportional to the distance from the source.
Note 2 to entry: The term "far field" is used in this document only in connection with non-focusing source transducers. For focusing transducers, a different terminology for the various parts of the transmitted field applies (see IEC 61828 [5]).
Note 3 to entry: If the shape of the transducer aperture produces several transition distances, the one furthest from the transducer is used.
[SOURCE: IEC 62127-1:2022, 3.30]
3.19
free field
sound field in a homogeneous and isotropic medium in which the effects of boundaries are negligible
[SOURCE: IEC 60565:2006, 3.13]
3.20
hydrophone
transducer that produces electric signals in response to pressure fluctuations in water
Note 1 to entry: A hydrophone is principally a passive device designed and built to respond to sound pressure.
Note 2 to entry: In some applications, a hydrophone is used as an active device to transmit sound.
[SOURCE: IEC 60050-801:2021 [6], 801-32-26]
3.21
hydrophone assembly
[SOURCE: IEC 62127-3:2022, 3.13]
3.22
hydrophone axis
nominal symmetry axis of the hydrophone active element
Note 1 to entry: Unless stated otherwise (explicitly and quantitatively) by the manufacturer, it is understood for the purposes of this document that this is given by the apparent geometrical symmetry axis of the hydrophone.
[SOURCE: IEC 62127-3:2022, 3.14]
3.23
hydrophone geometrical size
geometrical size of a hydrophone active element
ag
size defined by the lateral extents of the active element of a hydrophone
Note 1 to entry: The hydrophone geometrical size is expressed in metres (m).
Note 2 to entry: For hydrophones with a circular geometry, the hydrophone geometrical size is a radius.
Note 3 to entry: For hydrophones with a rectangular geometry, the hydrophone geometrical size is a half the greater value of the length or width.
Note 4 to entry: For fibre optic hydrophones, the hydrophone geometrical size is the radius of the light spot at the tip of the fibre.
[SOURCE: IEC 62127-3:2022, 3.11, modified - Note 4 to entry has been added.]
3.24
hydrophone pre-amplifier
active electronic device connected to, or intended to be connected to, a particular hydrophone and reducing its output impedance
Note 1 to entry: A hydrophone pre-amplifier requires a supply voltage (or supply voltages).
Note 2 to entry: The hydrophone pre-amplifier can have a forward voltage transmission factor of less than one, i.e. it need not necessarily be a voltage amplifier in the strict sense.
[SOURCE: IEC 62127-3:2022, 3.15]
3.25
instantaneous acoustic pressure
p (t)
pressure at a particular instant in time and at a particular point in an acoustic field, minus the ambient pressure
Note 1 to entry: Instantaneous acoustic pressure is expressed in pascals (Pa).
[SOURCE: IEC 60050-802:2011 [7], 802-01-03, modified - Note 1 to entry has been added.]
3.26
instantaneous intensity
I (t)
acoustic energy transmitted per unit time in the direction of acoustic wave propagation per unit area normal to this direction at a particular instant in time and at a particular point in an acoustic field
Note 1 to entry: Instantaneous intensity is the product of instantaneous acoustic pressure and particle velocity. It is difficult to measure intensity in the ultrasound frequency range. For the measurement purposes referred to in this document and under conditions of sufficient distance from the external transducer surface (at least one transducer diameter, or an equivalent transducer dimension in the case of a non-circular transducer) the instantaneous intensity can be approximated by the derived instantaneous intensity.
Note 2 to entry: Instantaneous intensity is expressed in units of watt per metre squared (W/m2).
[SOURCE: IEC 62127-1:2022, 3.36]
3.27
longitudinal plane
plane defined by the beam axis and a specified orthogonal axis
Note 1 to entry: See IEC 62127-1:2022, Figure 1.
[SOURCE: IEC 62127-1:2022, 3.39].]
3.28
near field
region of the field where mml-m9 z < z T aligned along the beam axis for planar non-focusing transducers
Note 1 to entry: For circular planar transducers, this is at a distance less than Aob/πλ, where Aob is the output beam area and λ is the wavelength of the ultrasound corresponding to the acoustic frequency.
Note 2 to entry: If the shape of the transducer aperture produces several transition distances, the one closest to the transducer is used.
[SOURCE: IEC 62127-1:2022, 3.41].]
3.29
reference centre
point on or near a hydrophone about which its acoustic receiving sensitivity is defined
Note 1 to entry: Unless stated otherwise (explicitly and quantitatively) by the manufacturer, it is understood for the purposes of this document that this is given by the geometrical centre of the front surface of the hydrophone active element.
[SOURCE: IEC 62127-3:2022, 3.16]
3.30
source aperture plane
closest possible measurement plane to the external transducer surface plane that is perpendicular to the beam axis
Note 1 to entry: If the offset distance is zero, the source aperture plane can be coincident with the external transducer surface plane.
Note 2 to entry: See Figure B.5, Figure B.7, Figure B.9 and Figure B.10 in IEC 61828:2020 [5].
[SOURCE: IEC 61828:2020 [5], 3.135].]
3.31
source aperture width
LSA
<in a specified longitudinal plane> greatest −20 dB beamwidth along the line of intersection between the designated longitudinal plane and the source aperture plane
Note 1 to entry: See Figure B.5 and Figure B.6b in IEC 61828:2020 [5].
Note 2 to entry: Source aperture width is expressed in metres (m).
[SOURCE: IEC 61828:2020 [5], 3.136, modified - two Notes to entry on HITU transducers have been omitted.].]
3.32
transducer aperture width
LTA
full width of the transducer aperture along a specified axis orthogonal to the beam axis of the unsteered beam at the centre of the transducer
Note 1 to entry: See Figure 3 in IEC 62127-1:2022.
Note 2 to entry: Transducer aperture width is expressed in metres (m).
[SOURCE: IEC 62127-1:2022, 3.81].]
3.33
transition distance
zT
aperture area divided by π times the effective wavelength λeff
[SOURCE: IEC 61828:2020 [5], definition 3.153, modified - In the definition, the words "the effective wavelength" have been added and a subscript has been added to the symbol for effective wavelength.]
3.33.1
transition distance for design
zTD
for a given longitudinal plane, the transducer aperture area ATA of the ultrasonic transducer divided by π times the effective wavelength λeff
mml-m10 z TD = A TA / ( π λ eff )
 (5)
Note 1 to entry: For design, for an unapodized ultrasonic transducer with circular symmetry about the beam axis, the source aperture area is mml-m11 π a e 2, where ae is the effective radius; therefore the transition distance is mml-m12 z T = a 2 / λ eff.
Note 2 to entry: For a spherically focusing transducer, the transition distance is approximately the same but more exactly it is mml-m13 z T = 2 D 2 ( 1 − 1 − ( a / D ) 2 ) / λ eff where D is the radius of curvature.
Note 3 to entry: For design, for an unapodized rectangular ultrasonic transducer which has a transducer aperture width, LTA1, in a specified longitudinal plane, the effective in-plane area is (LTA1)2. Therefore, for this plane, the transition distance is mml-m14 z T1 = ( L TA1 ) 2 / ( π λ eff ). The transition distance for the orthogonal longitudinal plane including the second transducer aperture width is mml-m15 z T2 = ( L TA2 ) 2 / ( π λ eff ).
Note 4 to entry: For apodized transducers with a symmetric even apodization or weighting function W (normalized to a maximum value of one for the particle velocity distribution), the transition distances are the following.
For circularly symmetric transducers with an active radius a: mml-m16 z T = ( 2 λ ) ∫ 0 a W ( r )   r d r
For a rectangular transducer with a physical aperture length L in a given longitudinal plane, for example, xz: mml-m17 z T = ( 4 π λ eff ) | ∫ 0 L / 2 W ( x )   d x | 2
Note 5 to entry: For symmetric shapes other than the most common cases of circular symmetry and rectangular geometry, the same definition of transition distance can be used. For example, for apertures with n-fold symmetry (n > 2) such as hexagons and octagons, the transition distance in a symmetry plane perpendicular to a side is equal to the area of the aperture divided by (πλeff). For annular arrays with several rings, the equivalent area is the total (all rings) area of the active aperture. For cases in which unique phasing is used such as annular arrays with alternate phase shifts (0° or 180°) in addition to intended focusing or for cases of apertures with unusual shapes and phasing, three-dimensional diffraction computation is employed to determine the minimum beamwidth corresponding to the appropriate transition distance.
Note 6 to entry: Transition distance for design is expressed in metres (m).
[SOURCE: IEC 61828:2020 [5], 3.153.1, modified - A subscript has been added to the symbol for effective wavelength and Note 6 to entry has been omitted.]
3.33.2
transition distance for measurement
zTM
for a given longitudinal plane, the source aperture area, ASAeff , of the ultrasonic transducer divided by π times the effective wavelength λeff
mml-m18 z TM = A SAeff / ( π λ eff )
 (6)
Note 1 to entry: For measurements in each specified longitudinal plane, the source aperture width in that plane is used, or mml-m19 z T1 = ( L SA1 ) 2 / ( π λ eff ); in the other orthogonal plane, mml-m20 z T2 = ( L SA2 ) 2 / ( π λ eff ).
Note 2 to entry: Transition distance for measurement is expressed in metres (m).
[SOURCE: IEC 61828:2020 [5], 3.153.2, modified - A subscript has been added to the symbol for effective wavelength.]
3.34
uncertainty
parameter, associated with the result of a measurement, that characterizes the dispersion of the values that could reasonably be attributed to the measurand
Note 1 to entry: See ISO/IEC Guide 98-3:2008 [10], 2.2.3.
3.35
wavelength
λ
distance in the direction of propagation of a periodic wave between two successive positions at which the phase is the same
Note 1 to entry: The wavelength is expressed in metres (m).
[SOURCE: IEC 60050-845:2020 [8], 845-21-025, modified - Notes 1, 2, 3, 5 and 6 to entry have been omitted. Note 4 to entry has been renumbered Note 1 to entry and "or decimal submultiples of metre" has been deleted.]
30mm
20mm
15mm
15mm
10mm
17mm

Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
Gulf StandardGSO IEC 62127-2:2026
GSO IEC 62127-2:2026Gulf Standard
Bibliography
[1]
HARRIS, GR., HOWARD SM., HURRELL, AM., LEWIN, PA., SCHAFER, ME., WEAR, KA., WILKENS, V. and ZEQIRI, B. Hydrophone measurement for biomedical ultrasound applications: a review, IEEE Trans. Ultrason., Ferroelect., Freq. Contr., Febr. 2023, vol. 70, iss. 2, p. 85-100
[2]
IEC 60565-1, Underwater acoustics - Hydrophones - Calibration of hydrophones - Part 1: Procedures for free-field calibration of hydrophones
[3]
IEC 60565-2, Underwater acoustics - Hydrophones - Calibration of hydrophones - Part 2: Procedures for low frequency pressure calibration
[4]
HARRIS, GR. Are current hydrophone low frequency response standards acceptable for measuring mechanical/cavitation indices? Ultrasonics, 1996, vol. 34, iss. 6, p. 649-654
[5]
IEC 61828:2020, Ultrasonics - Transducers - Definitions and measurement methods regarding focusing for the transmitted fields
[6]
IEC 60050-801, International Electrotechnical Vocabulary (IEV) - Part 801: Acoustics and electroacoustics
[7]
IEC 60050-802, International Electrotechnical Vocabulary (IEV) - Part 802: Ultrasonics
[8]
IEC 60050-845, International Electrotechnical Vocabulary (IEV) - Part 845: Lighting
[9]
RADULESCU, EG., LEWIN, PA., NOWICKI, A. and BERGER, WA. Hydrophones' effective diameter measurements as a quasi-continuous function of frequency. Ultrasonics, 2003, vol. 41, iss. 8, p. 635-641
[10]
ISO/IEC Guide 98-3:2008, Uncertainty of measurement - Guide to the expression of uncertainty in measurement (GUM:1995)
[11]
RAJAGOPAL, S., ROBINSON, S. P., ABLITT, J., MILORO, P., WANG, L., ZEQIRI, B., HURRELL, A. On the importance of consistent insonation conditions during hydrophone calibration and use," IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., Feb. 2023, vol. 70, iss. 2. p. 120-127
[12]
FAY, B., LEWIN, PA., LUDWIG, G., SESSLER, GM. and YANG, G. The influence of spatial polarization distribution on spot poled PVDF membrane hydrophone performance. Ultrasound Med. Biol., 1992, vol. 18, no. 6-7, p. 625-635
[13]
RADULESCU, EG., WÓJCIK, J., LEWIN, PA., NOWICKI, A. and BERGER, WA. The influence of finite aperture and frequency response of ultrasonic hydrophone probes on the determination of acoustic output. Ultrasonics, April 2004, vol. 42, iss. 1-9, p. 367-372
[14]
RADULESCU, EG., WÓJCIK, J., LEWIN, PA. and NOWICKI, A. Probing acoustic fields of clinically relevant transducers: the effect of hydrophone probes' finite apertures and bandwidths. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., October 2004, vol. 51, iss. 10, p. 1262-1270
[15]
HARRIS, GR., GAMMELL, PA., RADULESCU, EGR. and LEWIN, PA. Interlaboratory evaluation of hydrophone sensitivity calibration from 0.1 to 2 MHz via time delay spectrometry. Ultrasonics, April 2004, vol. 42 iss. 1-9, p. 349-353
[16]
ISO/IEC Guide 99:2007, International vocabulary of metrology - Basic and general concepts and associated terms (VIM)
[17]
PEDERSON, PC., LEWIN, PA. and BJORNO, L. Application of time-delay spectrometry for calibration of ultrasonic transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., March 1988, vol. 35, iss. 2, p. 185-205
[18]
GAMMELL, PM. and HARRIS, GR. Time delay spectrometry for hydrophone calibrations below 1 MHz. J. Acoust. Soc. Am., 1999, vol. 106, L41-6
[19]
ISAEV, A.E. and Matveev A.N. Calibration of Hydrophones in a Field with Continuous Radiation in a Reverberating Pool. Acoustical Physics, 2009, vol. 55, no. 6, p. 762-770
[20]
COSTA-FELIX, R.P.B. and MACHADO, J.C. Output bandwidth enhancement of a pulsed ultrasound system using a flat envelope and compensated frequency-modulated input signal: Theory and experimental applications. Measurement, 2015, vol. 69, p. 146-154
[21]
ISAEV, A.E. A Quality Criterion for Obtaining Free-Field Conditions when Calibrating a Hydroacoustic Receiver in a Water Tank with Reflecting Sides. Measurement Techniques, 2014, vol. 57, no. 5, p. 549-556
[22]
BRENDEL, K., LUDWIG, G. Calibration of Ultrasonic Standard Probe Transducers. Acustica, 1976, vol. 36, p. 203
[23]
SHOU W, DUAN S, HE P, XIA R, QIAN D. Calibration of a Focusing Transducer and Miniature Hydrophone As Well As Acoustic Power Measurement Based on Free-Field Reciprocity in a Spherically Focused Wave Field. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., 2006, vol. 53 no. 3, p. 564-570
[24]
OLIVEIRA, E.G., COSTA-FELIX, R.P.B. and MACHADO, J.C. Primary reciprocity-based method for calibration of hydrophone magnitude and phase sensitivity: complete tests at frequencies from 1 to 7 MHz. Ultrasonics, 2015, vol. 58, p. 87-95
[25]
BEISSNER, K. On the plane-wave approximation of acoustic intensity. J. Acoust. Soc. Am., 1982, vol. 71, p. 1406-1411
[26]
FICK, SE. The NIST power reference source. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry. Boca Raton: CRC Press, 1992, Ch. 6, p. 169-183
[27]
BEISSNER, K. Radiation force and force balances. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry. Boca Raton: CRC Press, 1992, p. 127-142
[28]
WILKENS, V. and MOLKENSTRUCK, W. Broadband PVDF membrane hydrophone for comparisons of calibration methods up to 140 MHz. IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 2007, vol. 54, p. 1784-1791
[29]
BEYER, RT. Nonlinear Acoustics. U.S. Government Printing Office, 1975
[30]
CARSTENSEN, EL. and MUIR, TG. The role of nonlinear acoustics in biomedical ultrasound. In GREENLEAF, J., ed., Ultrasound Tissue Characterization. Boca Raton: CRC Press, 1986, Ch. 3, p. 57-79
[31]
BJORNO, L. and LEWIN, PA. Measurement of B/A parameter in tissues. In GREENLEAF, J., ed., Ultrasound Tissue Characterization. Boca Raton: CRC Press, 1986, Ch. 6, p. 141-163
[32]
LEWIN, PA. Practical implementations and technology of measurement devices. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry: Boca Raton: CRC Press, 1992. Ch. 7, p. 185-215
[33]
DUCK, FA. and MARTIN, K. Exposure values for medical devices. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry. Boca Raton: CRC Press, 1992, Ch. 11, p. 315-344
[34]
BACON, DR. Characteristics of a PVDF membrane hydrophone for use in the range 1-100 MHz. IEEE Transactions on Sonics and Ultrasonics, January 1982, vol. SU-29, no. 1, p. 18-25
[35]
LEWIN, PA. Linearity of the polymer probes. Proc. WFUMB. Sydney. Elmsford, NY: Pergamon Press, 1985, p. 537
[36]
KRAYNAK, T., SCHAFER, ME. and KRAKHMAN, V. Development of a cost-effective shock wave hydrophone. 1994 IEEE Ultrasonics Symposium Proceedings, 1994, p. 1805-1808
[37]
DODICK, JM. Surgical instrument with input power transducer. US Patent No. 5324282
[38]
LEWIN, PA., BHATIA, R., ZHANG, Q., and DODICK, J. Characterization of optoacoustic surgical devices. IEEE Ultrasonics Symposium, 1995
[39]
SCHAFER, ME. Techniques of hydrophone calibration. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry. Boca Raton: CRC Press, 1992, Ch. 8, p. 217-255
[40]
TRIER, HG. Ultrasonic devices for surgery (cataract removal and viterectomy) in ophthalmology. J. d'Echographie Med. Ultrasonore, 1985, vol. 1, p. 17-23
[41]
BEISSNER, K. Schallfelduntersuchungen an ophthalmologisch-chirurgischen Ultraschallgeraeten. In Fortschritte d. Akustik - DAGA '80. Berlin: VDE-Verlag, 1980, p. 567-570
[42]
SCHAFER, ME. and BROADWIN, A. Acoustical characterization of ultrasonic surgical devices. 1994 IEEE Ultrasonics Symposium Proceedings, 1994, p. 1903-1906
[43]
MEEKS, S. and TING, R. The evaluation of static and dynamic stress on the piezoelectric and dielectric properties of PVDF. J. Acoust. Soc. Am., 1984, vol. 75, p. 1010
[44]
TANCRELL, RH., WILSON, DT. and RICKETTS, D. Properties of PVDF polymer for sonar. 1985 IEEE Ultrasonics Symposium Proceedings, 1985, p. 624-627
[45]
HERMAN, BA. and HARRIS, GR. Calibration of miniature ultrasonic receivers using a planar scanning technique. J. Acoust. Soc. Am., 1982, vol. 72, p. 1357-1363
[46]
PINKERTON, JMM. The absorption of ultrasonic waves and liquids and its relation to molecular constitution. Proc. Phys. Soc., 1949, vol. 62, p. 129-141
[47]
CORBETT, SS. The influence of nonlinear fields on miniature hydrophone calibration using the planar scanning technique. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., March 1988, vol. 35, iss. 2, p. 162-167
[48]
BEISSNER, K. Maximum hydrophone size in ultrasonic field measurements. Acustica, 1985, vol. 59, p. 61-66
[49]
RADULESCU, EG., LEWIN, PA., GOLDSTEIN, A. and NOWICKI, A. Hydrophone spatial averaging corrections from 1-40 MHz. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., November 2001, vol. 48, iss. 6, p. 1575-1580
[50]
RADULESCU, EG., LEWIN, PA. and NOWICKI, A. 1-60 MHz Measurements in Focused Acoustic Fields using Spatial Averaging Corrections. Ultrasonics, May 2002, vol. 40, iss. 1-8, p. 497-501
[51]
AIUM/NEMA. Safety standard for diagnostic ultrasound equipment. J. Ultrasound Med., 1983, Apr, 2 (4 Suppl), S1-50. PMID: 6854728.
[52]
BACON, DR., A new method for ultrasonic hydrophone calibration. 1982 IEEE Ultrasonics Symposium Proceedings, 1982, p. 700-704
[53]
LOCKWOOD, JC., MUIR, TG. and BLACKSTOCK, DT. Directive harmonic generation in the radiation field of a circular piston. J. Acoust. Soc. Am., 1973, vol. 53, p. 1148-1153
[54]
DEL GROSSO, VA. and MADER, CW. Speed of sound in pure water. J. Acoust. Soc. Am., 1972, vol. 52, p. 1442-1446
[55]
BILANIUK, N. and WONG, GSK. Speed of sound in pure water as a function of temperature. J. Acoust. Soc. Am., 1993, vol. 93, p.2306
[56]
IEC TR 62781, Ultrasonics - Conditioning of water for ultrasonic measurements
[57]
RADULESCU, EG., WÓJCIK, J., LEWIN, PA. and NOWICKI, A. Nonlinear propagation model for ultrasound hydrophones calibration in the frequency range up to 100 MHz. Ultrasonics, June 2003, vol. 41, iss. 4, p. 239-245
[58]
RADULESCU, EG., LEWIN, PA., WÓJCIK, J. and NOWICKI, A. Calibration of ultrasonic hydrophone probes up to 100 MHz using time gating frequency analysis and finite amplitude waves. Ultrasonics, June 2003, vol. 41, iss. 4, p. 247-254
[59]
BLEEKER, HJ. and LEWIN, PA. A new method of ultrasonic hydrophone calibration using KZK wave modeling. J. Acoust. Soc. Am., 1998, vol. 103 p. 2962
[60]
BLEEKER, HJ. and LEWIN, PA. A novel method for determining calibration and behavior of PVDF ultrasonic hydrophone probes in the frequency range up to 100 MHz. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., November 2000, vol. 47, iss. 6, p. 1354-1362
[61]
LEWIN, PA., UMCHID, S., SUTIN, A. and SARVAZYAN, A. Beyond 40 MHz frontier: the future technologies for calibration and sensing of acoustic fields. J. Phys.: Conf. Ser., 2004, vol. 1, p. 38-43
[62]
REIBOLD, R. and MOLKENSTRUCK, W. Laser interferometric measurement and computerized evaluation of ultrasonic displacements. Acta Acustica united with Acustica, 1981, vol. 49, iss.3, p. 205-211
[63]
REIBOLD, R. and MOLKENSTRUCK, W. Investigation of pulse-excited hydrophones for ultrasonic field measurements using laser interferometry. Ultrasonics, March 1987, vol. 25, iss. 2, p. 114-118
[64]
SCRUBY, CB. and DRAIN, LE. Laser Ultrasonics: Techniques and Applications. Bristol: Adam Hilger, 1990
[65]
REIBOLD, R. and MOLKENSTRUCK, W. Optical measuring techniques. In ZISKIN, MC. and LEWIN, PA., eds., Ultrasonic Exposimetry. Boca Raton: CRC Press, 1992, Ch. 5, Part C, p. 143-162
[66]
KOCH, Ch., LUDWIG, G. and MOLKENSTRUCK, W. Calibration of an interferometric fiber tip sensor for ultrasound detection. Ultrasonics, June 1997, vol. 35, iss. 4, p. 297-303
[67]
KOCH, C. and MOLKENSTRUCK, W. Primary calibration of hydrophones with extended frequency range 1 to 70 MHz using optical interferometry. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., September 1999, vol. 46, iss. 5, p. 1303-1314
[68]
WEBER, M. and WILKENS, V. Using a heterodyne vibrometer in combination with pulse excitation for primary calibration of hydrophones in amplitude and phase. Metrologia, 2017, vol. 54, p. 432-444
[69]
WEBER, M. and WILKENS, V. A comparison of different calibration techniques for hydrophones used in medical ultrasonic field measurement. IEEE Trans. Ultrason., Ferroelect., Freq. Contr. 2021, vol. 68, no.5, p. 1919-1929
[70]
COOLING, M. P., HUMPHREY, V. F., and WILKENS, V. Hydrophone area-averaging correction factors in nonlinearly generated ultrasonic beams. J. Phys. Conf. Ser. 279, Adv. Metrol. For Ultrasound in Med., 2011, 012002
[71]
REMBE, C., BODECKER, S., DRABENSTEDT, A., PUDEWILLS, F. and SIEGMUND, G. Heterodyne laser-Doppler vibrometer with a slow-shear-mode Bragg cell for vibration measurements up to 1.2 GHz. 8th Int. Conf. on Vibration Measurements by Laser Techniques, 2008, 7098, 70980A
[72]
WILKENS, V. and KOCH, C. Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone. J. Acoust. Soc. Am., 2004, vol. 115, p. 2892-2903
[73]
WINTER, M., FUSER, H., BIELER, M., SIEGMUND, G. and REMBE, C. The problem of calibrating laser-Doppler vibrometers at high frequencies. AIP Conf. Proc., 2012, vol. 1457, p. 165-175
[74]
REMBE, C. and DRABENSTEDT, A. Laser-scanning confocal vibrometer microscope: theory and experiments. Rev. Sci. Instrum., 2006, vol. 77, 083702
[75]
BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP, and OIML Evaluation of Measurement Data - Supplement 2 to the Guide to the Expression of Uncertainty in Measurement Extension to Any Number of Output Quantities, 2011, vol 102 (Sèvres: Joint Commitee for Guides in Metrology (JCGM) Bureau International des Poids et Mesures (BIPM))
[76]
RAJAGOPAL, S., FURY, C.R., ZEQIRI, B., BRANDT, M., WILKENS, V., KOCH, C., MATSUDA, Y., YOSHIOKA, M., PING, Y., YAN, Z., WENPING, B. COSTA-FELIX, R.P.B., OLIVEIRA, E.G. Report on BIPM/CIPM key comparison CCAUV.U-K4: absolute calibration of medical hydrophones in the frequency range 0.5 MHz to 20 MHz, Metrologia, 2016, vol. 53, 09004
[77]
RAJAGOPAL, S. and COX, T. 100 MHz bandwidth planar laser-generated ultrasound source for hydrophone calibration. Ultrasonics, 2020, vol. 108, 106218
[78]
BACON, DR. Primary calibration of ultrasonic hydrophone using optical interferometry. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr. March 1988, vol. 35, iss. 2, p. 152-161
[79]
YANG, P., XING, G., and HE, L. Calibration of high-frequency hydrophone up to 40 MHz by heterodyne interferometer, Ultrasonics, 2014, vol. 54, no. 1, p. 402-407
[80]
MATSUDA, Y., YOSHIOKA, M., and UCHIDA , T., Absolute hydrophone calibration to 40 MHz using ultrasonic far-field, Mater. Trans., 2014, vol. 55, no. 7, p. 1030-1033
[81]
BACON, DR. and ROBINSON, SP. Intercomparison of 1 mm hydrophone calibrations in the frequency range 0.5 to 15 MHz. (Report EUR 13525 EN). Bureau Communautaire de Référence, Commission of the European Communities, 1991
[82]
ROBINSON, SP., BACON, DR. and MOSS, BC. The measurement of the frequency response of a photodiode and amplifier using an opto-mechanical frequency response calibrator. Meas. Sci. Technol., 1990, vol. 1, p. 1184-1187
[83]
ESWARD, TJ. and ROBINSON, SP. Extending the frequency range of the National Physical Laboratory primary standard laser interferometer for hydrophone calibrations to 80 MHz. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., May 1999, vol. 46, iss. 3, p. 737-744
[84]
PRESTON, RC., BACON, DR., LIVETT, AJ. and RAJENDRAN, K. PVDF membrane hydrophone performance properties and their relevance to the measurement of the acoustic output of medical ultrasonic equipment. J. Phys. E: Sci. Instrum., 1983, vol. 16, p. 786-796
[85]
SMITH, RA. and BACON, DR. A multiple-frequency hydrophone calibration technique. J. Acoust. Soc. Am., 1990, vol. 87, p. 2231-2243
[86]
BACON, DR. Nonlinear acoustics in ultrasound calibration and standards. In HAMILTON, MF. and BLACKSTOCK, DT., eds., Frontiers of nonlinear acoustics. 12th ISNA. Elsevier, 1990, p. 3-19
[87]
HARRIS, GR. and SHOMBERT, DG. A pulsed near-field technique for measuring the directional characteristics of acoustic receivers. IEEE Transactions on Sonics and Ultrasonics, 1985, vol. 32, no. 6, p. 802-808
[88]
LEWIN, PA. Calibration and performance evaluation of miniature ultrasonic hydrophones using time delay spectrometry. 1981 IEEE Ultrasonics Symposium Proceedings, 1981, p. 660-664
[89]
LUDWIG, G. and BRENDEL, K. Calibration of hydrophones based on reciprocity and time delay spectrometry. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., March 1988, vol. 35, iss. 2, p. 168-174
[90]
WILKENS, V., SONNTAG, S. and WEBER, M. Secondary complex-valued hydrophone calibration up to 100 MHz using broadband pulse excitation and a reference membrane hydrophone. Proc. 2019 IEEE Ultrasonics Symposium, 2019, Glasgow, Scotland, (2019), WeH5.2
[91]
SELFRIDGE, A. and LEWIN, PA. Wideband Spherically Focused PVDF Acoustic Sources for Calibration of Ultrasound Hydrophone Probes. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., November 2000, vol. 47, iss. 6, p. 1372-1376
[92]
HARRIS, GR., CAROME, EF. and DARDY, HD. An analysis of pulsed ultrasonic fields as measured by PVDF spot-poled membrane hydrophones. IEEE Transactions on Sonics and Ultrasonics, September 1983, vol. SU-30, no. 5, p. 295-303
[93]
BABOUX, JC., LAKESTANI, F. and PERDRIX, M. Theoretical and experimental study of the contribution of radial modes to the pulsed ultrasonic field radiated by a thick piezoelectric disk. J. Acoust. Soc. Am., 1984, vol. 75, p. 1722-1731
[94]
HEYSER, RC. Acoustical measurements by time delay spectrometry. J. Audio Eng. Soc., 1967, vol. 15, p. 370
[95]
CHIVERS, RC. Time-delay spectrometry for ultrasonic transducer characterization. J. Phys. E. Sci. Instrum., 1986, vol. 19, p. 834-843
[96]
KOCH, C. Amplitude and Phase Calibration of Hydrophones by Heterodyne and Time-gated Time-delay Spectrometry. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., March 2003, vol. 50, iss. 3, p. 344-348
[97]
GAMMELL, PM. Time and frequency domain measurements of materials with high ultrasonic attenuation using time domain spectroscopy. In THOMPSON, DO. and CHIMENTI, DE., eds, Review of Progress in Quantitative NDE. Plenum Press, 1986, vol. 5, p. 759-656
[98]
LEWIN, PA. Miniature piezoelectric polymer ultrasonic hydrophone probes. Ultrasonics, September 1981, vol. 19, iss. 5, p. 213-216
[99]
LEWIN, PA. and SCHAFER, ME. Ultrasonic probes in measurement practice. Medical Device and Diagnostic Industry, 1986, vol. 8, no. 5, p. 40-45
[100]
HURRELL, A. Voltage to pressure conversion: Are you getting 'phased' by the problem. J. Phys. Conf. Ser., 2004, vol.1, iss. 1, p. 57-62
[101]
WEAR K., GAMMELL P., MARUVADA S., LIU Y., and HARRIS G. Time-delay spectrometry measurement of magnitude and phase of hydrophone response. IEEE Trans UFFC. 11/2011; 58(11):2325-33
[102]
BLOOMFIELD P., GANDHI G., and LEWIN P. Membrane hydrophone phase characteristics through nonlinear acoustics measurements. IEEE Trans UFFC. 11/2011; 58(11):2418-37
[103]
BLOOMFIELD P., GANDHI G., and LEWIN P. Nonlinear acoustics determination of phase characteristics of PVDF membrane hydrophones. J. Phys: Conf. Ser. 279 012001
[104]
WILKENS, V. Characterization of an optical multilayer hydrophone with constant frequency response in the range from 1 to 75 MHz. J. Acoust. Soc. Am., 2003, vol. 113, p. 1431-38
[105]
WILKENS, V. and KOCH, C. Optical multilayer detection array for fast ultrasonic field mapping. Opt. Lett., 1999, vol. 24, no. 15, p. 1026-1028
[106]
EICHSTAEDT, S. and WILKENS, V. GUM2DFT - A software tool for uncertainty evaluation of transient signals in the frequency domain. Meas. Sci. Technol., 2016, vol. 27, no. 5, 055001-1-055001-12
[107]
HUMPHREY, VF., COOLING, MP., DUNCAN, TM. and DUCK, F. Measurement of the phase response of a membrane hydrophone and its application to ultrasonic field characterisation. Advanced Metrology for Ultrasound in Medicine (AMUM) conference, Teddington, UK, 27-28 April 2004
[108]
HUMPHREY, VF., COOLING, MP. A nonlinear propagation model-based phase calibration technique for membrane hydrophones. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr., January 2008, vol. 55, iss. 1, p. 84-93
[109]
ZEQIRI, B. and BOND, AD. The influence of waveform distortion on hydrophone spatial-averaging corrections - Theory and measurement. J. Acoust. Soc. Am., 1992, vol. 92, p. 1809-1821
[110]
DU, G. and BREAZEALE, MA. Harmonic distortion of a finite amplitude Gaussian beam in a fluid. J. Acoust. Soc. Am., 1986, vol. 80, p. 212-216
[111]
BRENDEL, K., LUDWIG, G. Measurement of Ultrasonic Diffraction Loss for Circular Transducers. Acustica, 1975, vol. 32, p.110
[112]
BEISSNER, K. Free-field Reciprocity Calibration in the Transition Range between Near Field and Far Field. Acustica, 1980, vol. 46, p.162
[113]
FAY, B. Numerische Berechnung der Beugungsverluste im Schallfeld von Ultraschallwandlern. Acustica, 1976, vol. 36, p. 209
[114]
KHIMUNIN, A.S. Numerical Calculation of the Diffraction Corrections for the Precise Measurement of Ultrasound Absorption. Acustica, 1972, vol. 27, p. 173
[115]
BEISSNER, K. Exact integral expression for the diffraction loss of a circular piston source. Acustica, 1981, vol. 49, p. 212
[116]
YOSHIOKA, SATO, KIKUCHI and MATSUDA. Influence of ultrasonic nonlinear propagation on hydrophone calibration using two-transducer reciprocity method. Jpn. J. App. Phys., 2006, vol. 45, No.5B, p.4547-4549
[117]
PINKERTON, J.M.M. A Pulse Method for the Measurement of Ultrasonic Absorption in Liquids Results for Water. Nature, 1947, vol. 160, p. 128
[118]
BILANIUK, N, WONG, G.S.K. Erratum: Speed of sound in pure water as a function of temperature, J. Acoust. Soc. Am., 1996, vol. 99, no. 5, p. 3257
[119]
LUBBERS, J, GRAAFF, R. A simple and accurate formula for the sound velocity in water, Ultrasound Med. Biol., 1998, vol 24, no. 7, p. 1065-1068
[120]
BELOGOL'SKII, V.A., SEKOYAN, S.S., SAMORUKOVA, L.M. STEFANOV, S.R., Levtsov, V.I. Pressure dependence of the sound velocity in distilled water, Measurement Techniques, 1999, vol. 42, no. 4, p. 406-413
[121]
IEC TS 62903, Ultrasonics - Measurements of electroacoustical parameters and acoustic output power of spherically curved transducers using the self-reciprocity method
[122]
SHOU, W., DUAN, S., HE, P., XIA, R. and QIAN, D. Calibration of a focusing transducer and miniature hydrophone as well as acoustic power measurement based on free-field reciprocity in a spherically focused field, IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 2006, vol. 53, no.3, p. 564-570
[123]
LONGYANG, J., SHOU, W. and HU, B. Reciprocity calibration of hydrophones in high intensity focused ultrasound field, American Journal of Physics and Applications, 2020, vol. 8, no.2, p. 19-24
30mm
20mm
15mm
15mm
10mm
17mm

1Numbers in square brackets refer to the Bibliography.